Defibrillator / Transcutaneous Pacemaker Analyzer

Product Code : Defibrilator Transcutaneous Pacemaker analyzer

The DELTA 3300 Defibrillator / Transcutaneous Pacemaker Analyzer is a precision instrument for testing defibrillators and external (Transcutaneous) Pacemakers. It combines the complete testing functions of a Defibrillator Tester, Transcutaneous Pacemaker Analyzer, and 12-Lead ECG / Arrhythmia Simulator into one compact, lightweight, and easy-to-use instrument.

The Delta 3300 tests all Defibrillators including AEDs. It measures both monophasic and Bi-Phasic waveforms.  It is compact, lightweight and easy to use. The test results can be saved in the internal memory and uploaded to the PC and printed using ProComPrint software (CD included with the unit).

Show More... Add to cart

    PRODUCT HIGHLIGHTS
     
    • Energy and cardioversion measurement
    • Peak voltage and current reading
    • Waveform Storage and Playback
    • Tests all Automated defibrillators
    • Pacemaker loads from 50 to 1100 Ohm
    • Pacer Amplitude, Rate, Pulse width
    • Demand Sensitivity, Refractory period, immunity and pulse Energy
    • 12 Lead ECG, Arrhythmia, Performance
    • Li-Ion Rechargeable Battery

     

    FEATURES
    • Large LCD Display with Backlight
    • Simple and easy to use
    • Compact & Lightweight (3lbs)
    • Test results can be saved & uploaded to PC using Netech ProComPrint software
    • Li-Ion Rechargeable Battery
    • Hard Carrying case
    • 2 Year Warranty
    • Incredible Value

     

    DEFIBRILLATOR ANALYZER  
       
    Energy Output Measurement:  
    Maximum Energy: 800 J
    Display Resolution: Based on Auto Range
    Auto Range: 2 - 9.99 J, 10-99.9 J, 100-800 J
    Resolution: 0.01 J ( 2 -9.99), 0.1 J (10-99.9), 1 J (100-500)
    Accuracy: ±1 % of reading ± 0.1 J (2-9.9), ± 1J (10-99), ± 2 % (100-500) ± 1J
    Load Resistance: 50 ohms ±1 %; non-inductive (< 1 μH)
    Sampling Time: 20 uS
    Measurement Time Window: 60 ms
    Absolute Max Peak Voltage: 5000 V
    Absolute Max Peak Current: 100 A
    Pulse Width: 60 ms
    High-Level Output:  
    High Range: 100:1 amplitude attenuation
    Low Range: 200:1 amplitude attenuation
    Waveform Storage / Playback: Discharge can be viewed via ECG outputs and paddles
       
    Sync Time Measurements:  
    Timing Window: Starts: 40 ms at each R wave peak
    Test Waveforms: All Waveform simulations available
    Delay Time Accuracy: ±1 ms
       
    Charge Time Measurement: 30 to 60 in steps of 15 seconds

     

    TRANSCUTANEOUS PACEMAKER ANALYZER
    Test Load Range: 50 -1100 Ohms in steps of 100 Ohms.
    Accuracy: 50 - 500 ohm ±1 %,  500 - 1100 ohm ±1.5 %
    Pulse Measurements:  
    Amplitude: 4 to 400 mA (50 ohm load)
    Accuracy: ±1 % or ± 1 mA
    Max. amplitude: 400 mA (50 Ohm Load)
    Rate: 30 to 800 ppm
    Accuracy: ±1 % or 2 ppm
    Pulse Width: 0.5 to 50 ms
    Accuracy: ±1 % or ±0.3 ms
    Demand Sensitivity Test:  
    Waveforms: QRS, Square(SQR), Triangle(TRI), and Haversine (SSQ)
    Waveform width: 40, 50, 60, 70, 80, 90, 100, 200, 300, 400 ms
    Pacer Rate: 30 to 120 ppm
    ECG Output:  
    Amplitude: User Selectable 0.1 - 5 mV
    Pacer Input (Load dependent):  
    Amplitude (50 ohm): 0 - 10 mV
    Resolution (50 ohm): 25 μV
    Amplitude: (500 ohm): 0 - 100 mV
    Resolution: (500 ohm): 1 mV
    Accuracy: ±2 %
    Immunity Test:  
    Interference Signal: 50/60 Hz
    ECG signal Amplitude: 0.5 to 5.0 mV
    Refractory Period Measurement: 20 to 500 ms (both Pacing and Sensing)
    Accuracy: ±5 ms

     

    PATIENT SIMULATOR
    ECG General:  
    Lead Configuration: 12-lead simulation RL, RA, LA, LL, V1-6
    Output Impedance: RL to All leads 1000 Ohms, V Leads: 1000 ohms to RL
    High Level Output: 0.5 V/mV of low level (lead II)
    Rates: 30, 40, 50, 60, 70, 80, 90, 100, 110, 120,130, 140, 150, 160, 170, 180, 190, 200, 220, 240, 260, 280, 300 & 310 BPM
    Accuracy: ±1 %
    Amplitude: User Selectable 0.1 - 5 mV (Lead II)
    Accuracy: ±2 % @ Lead II (SQR)
    Arrhythmia Selections: Ventricular Fibrillation Coarse (VFBC)
      Ventricular Fibrillation Fine (VFBF)
      Ventricular Tachycardia (VTAC)
      Atrial Fibrillation (AFIB)
    Performance Waveforms: Sine, Square, Triangle
    Rate: 0.2, 0.3, 0.4, 0.5, 1, 2, 3, 4, 5, 6, 7, 8, 9, 10, 15, 20, 25, 30, 35, 40, 45, 50, 55, 60 Hz.
    Amplitude: 0.1, 0.25, 0.50, 1, 1.5, 2.0, 2.5, 3.0, 3.5, 4.0, 4.5, 5.0 mV.
    Amplitude accuracy: ±2 % (Square Wave -Lead II 1.0 mV )
    Pacer Pulse:  
    Pulse Width: 10mS
    Amplitude: 1.0 mV (Lead II)
    Accuracy: ± 2 %

     

    GENERAL INFORMATION
    Temperature Range:  
    Operating: +15°C to +35°C
    Storage: 0°C to +50°C
    Display:  
    Type LCD with LED backlight
    Data Input/Outputs : RS-232C (1) for Computer control
    Power: 7.4 Volt Li-Ion Battery 100-230 VAC input, 850 mA DC Charger
    Housing: High impact ABS Plastic case
    Dimensions (D x W x H): 10.25 x 6.25 x 2.5 Inch, 26 X 16 x 6.4 Cm
    Weight: 1.5 kg / 3.3 lbs

     

Download Brochure Download Manual
Part Number Description Price
660 _Delta 3300 Defibrillator / Transcutaneous Pacemaker Analyzer Add to cart
660-152345 USB / Rs232 interface cable Add to cart
660-610-6383 Transcutaneous /AED Interface Cable (1 Pair Red / Black) Add to cart
660-Pro-Print ProCom Print software Add to cart
660-908 Interface cable for Medtronic Pacemakers Add to cart
660-651-HP Interface Cable for HP/ Philips Add to cart
660-???? For all other Interface Cables Please contact Netech Support Add to cart
660-CD-Manual Delta 3300 User manual CD Add to cart
660-PADDLE-ADAP Defibrillator Paddle Adapter ( 2 Pieces) Add to cart
660-HARD-CASE Defibrillator Paddle Adapter ( 2 Pieces) Add to cart
660-651-ZOLL Interface Cable for Zoll Medical Add to cart