Patient Simulators
MiniSim 1000-Basic - MiniSim Basic
|
The MiniSim 1000, Arrhythmia Simulator is a powerful,
comprehensive but basic patient simulator in a compact case.
It is designed for testing the performance of basic
patient monitoring instrumentation quickly and easily.
The microprocessor based instrument is menu driven via its
tactile feel keypad and easy to follow LCD display. The
MiniSim 1000 provides full 12 lead ECG simulation with 14
user selectabel rates from 30 to 350 BPM and 14 user
selectable amplitudes from 0.15 to 5 mV. Sine, Square, and
Triangle performance waveforms are simulated with 14 user
selectable rates from 0.1 to 100 Hz.The MiniSim also simulates
45 different arrhythmias including atrial, ventricular,
tall T wave, ST elevation, ST depression, myocardial infraction,
blocks, pulse, and pacer waveforms.
|
|
|
 |
 |
 |
12 Lead ECG |
 |
45 Arrhythmias |
 |
ECG Performance Waveforms |
 |
Easy to Use |
 |
Small, Compact Size |
 |
Menu driven operatoin |
 |
Large LCD Display |
 |
Tactile Keypad |
 |
Battery or AC Operation |
|

|
|
ECG: 12 leads with independent outputs referenced to RL.
NORMAL SINUS RHYTHM (NSR)
ECG Rates: 30, 60, 70, 80, 90, 100, 120, 150, 180, 210, 240, 270, 300, and 350 BPM. Accuracy 0.5%.
Amplitudes:
0.15, 0.3, 0.5, 1.0, 2.0, 3.0, 4.0, and 5.0 mV on Lead II. Lead 1 is
0.6 X Lead II, Lead III is 0.4 X Lead II, and Lead V is the same as
Lead II.
High Level: 500 X low level output on Lead II.
Accuracy: 2% (1-5mV).
PERFORMANCE WAVEFORMS (PERF)
Sine, Square, Triangle, Pulse
Frequencies: 0.1 to 0.9 in 0.1 Hz steps.
1.0 to 9.0 in 1.0 Hz steps.
10 to 100 in 10 Hz steps.
Accuracy: 1%
Amplitudes: 0.1, 0.2, 0.5, 1.0, 2.0, 3.0, 4.0, and 5.0 mV on Lead II.
Pulse: 20 ms pulse of 1mV amplitude repeated at 4 second intervals.
R WAVE DETECTION (RWD)
Widths: 10, 40, 50, 60, 70, 80, 90, 100, 110, and 120 ms.
Amplitudes: 0.15, 0.3, 0.5, 1.0, 2.0, 3.0, 4.0, 5.0 mV on Lead II.
PACER (PCR)
AP: Atrial Pacer set at 70 BPM.
ASP: Asynchronous Pacer - Ineffective Pacing.
Pacer Amplitudes: + 2 mV and - 2mV.
QRS Amplitudes: 1 mV
PCR: Pacer pulses alone.
Pacer Widths: 0.1, 0.2, 0.5, 1.0, 1.5, and 2.0 ms.
Amplitudes: 0.15, 0.3, 0.5, 1.0, 2.0, 3.0, 4.0, and 5.0 mV.
VP: Ventricular Pacer set at 70 BPM.
AVP: Atrial Ventricular Pacer set at 70 BPM.
QRS Amplitude: 1 mV.
Pacer Amplitude: -2mV.
|
POWER REQUIRMENTS:
One 9 Volt alkaline battery or optional AC adapter.
PHYSICAL CHARACTERISTICS:
Dimensions: 5.5 X 3.5 X 1.5 inches (13.9 X 8.9 X 3.8 cm).
Weight: 10 oz ( 0.3 kg).
TEMPERATURE REQUIREMENTS:
Operating: 59 to 95 Degrees F (15 to 35 Degrees C).
Storage: 32 to 131 Degrees F (0 to 55 Degrees C).
|
|
|
 |
|

 |