DELTA 3000 - Defibrillator and External Pacer Analyzer
 |
The
DELTA 3000 combines the complete testing functions of a defibrillator
tester, a transthoracic pacemaker analyzer, a transvenous pacemaker
analyzer and a 12 Lead ECG / Arrhythmia simulator into one compact,
lightweight, and easy to use instrument. No other instrument delivers
the performance testing power of the DELTA 3000.
The DELTA 3000
tests all defibrillators including AEDs. It measures both monophasic
and bi-phasic waveforms. It tests all external pacemakers including
transthoracic, transvenous, and AV sequential models. It has 12
built-in user selectable test loads for transthoracic pacing testing.
The built-in ECG / Arrhythmia simulator allows performance testing of
any ECG equipment.
The DELTA 3000 is operated via easy to follow
menus and eight soft keys. Instructions and test measurements are
displayed on the large four line LCD display. A print key outputs test
measurements to the RS 232 serial port.
The DELTA 3000 is
delivered in a hard carrying case with various paddle adapters and
interface cables. It provides unsurpassed biomedical testing
capabilites in one instrument.
|
|
|
 |
 |
 |
Compact, Lightweight, and Rugged |
 |
Simple, Menu Driven Operation |
 |
Large 4-line LCD |
 |
Simultaneous Display of All Test Measurements |
 |
Dedicated Print Key for Immediate Documentaion |
 |
RS 232 Interface |
 |
Battery or AC Operation |
 |
Two-Year Warranty |
|
 |
 |
Measures: Energy, Peak Voltage, Peak Current, and Cardio Sync. Time |
 |
Storage and Playback of output waveforms |
 |
Tests all External Pacemakers - Transthoracic, Transvenous, and AV Pacers |
 |
Measures: Amplitude, Rate, Pulse Width, Pulse Energy, Sensed and Paced Refactory Periods |
 |
50 and 60 Hz Immunity Tests |
 |
Transthoracic Test Loads: Fixed 50 Ohm or 12 User Selectable Loads to 1000 ohms |
 |
A / V Pacer Test Load: 500 Ohm |
 |
Tests all ECG Equipment. |
 |
12 lead ECG, User Selectabel Rates and Amplitudes |
 |
Performance Waveforms: Sine, Square, Triangle amp; Pulse |
 |
12 Arrhythmias |
|

|
|
Defibrillator Analyzer
Energy Measurement:
Load: 50 Ohm +/- 1 %
Range (High): 900 J Resolution: 1 Joule
Range (Low): 0 - 100 Resolution: 0.1 Joules
Accuracy: +/- 2% of the reading + 1 J for Hi Range or + 0.5 J for Low Range
Maximum Voltage: 5200 V
Maximum Current: 100 Amp (H)
Sync. Time: 0 - 250 ms
Waveform Output:
Real Time: Output 1000 : 1 Stored.
Discharge waveform expanded 200:1 and is repeated 8 sec. It can be viewed through the paddles and through the ECG High Level Output
Pacemaker Analyzer
Transcutaneous:
Test Load: 50 Ohm (Fixed) and variable 100 to 1000 Ohms user selectable
Current: 1 to 200 mA Accuracy: +/- 5% + 0.5 mA
Pulse Width: 0.5 to 80 ms Accuracy: +/- 1% + 0.1 ms
Refractory: 20 - 500 ms Accuracy: +/- 2% or + 10 ms
Atrial and Ventricular:
Test Load: 500 Ohms
Current: 1 to 25 mA Accuracy: +/- 5% + 0.5 mA
Pulse Rate: 30 to 800 ppm. Accuracy: +/- 1% + 0.1 ms
Refactory: +/- 20 - 500 ms
Accuracy: +/- 2% or 10 ms
Immunity Test: 50 / 60 Hz
ECG ArrhythmiaSimulator
ECG: Normal Sinus Rhythm.
Rate: 30, 60, 90, 120, 240
Performance Waveforms:
Sine, Square, Triangle and Pulse
Rate: 0.5, 1, 2, 10, 15, 20, 25, 40 Hz
Arrhythmias:
VFBC, VFBF, VTAC, AFIB, PVC, PVC1, PVC2, BGY, RBB, AFIB, ATRFT, NSR
Waveform Output:
12 Lead ECG and Paddle and High Level Output Jacks.

|
Power Requirements:
Two 9V Alkaline Batteries Or AC Adapter
 
Dimension:
10 x 11 x 2.25 inches (25.4 x 27.9 x 5.7 cm)
Weight:
4 lbs (1.8 kg) |
|
|
 |
|

 |